Home > Instruments & Meters > Optical Lens & Instrument > Spectrometer > DW-FP640 LCD Display Lab Equipment of Flame Photometer

DW-FP640 LCD Display Lab Equipment of Flame Photometer

DW-FP640 LCD Display Lab Equipment of Flame Photometer photo-1
US$ 440 - 5,000 MOQ: 1 Piece
Key Specifications
Get Latest Price
Place of Origin:
Shanghai, China
Brand Name:
DRAWELL
Model Number:
DW-FP640
Payment & Shipping
Payment Methods:
Port of Shipment:
Shanghai
Delivery Detail:
Delivery time depends on order quantity.
Place of Origin Shanghai, China
Brand Name DRAWELL
Model Number DW-FP640

DW-FP640 Flame Photometer

 

Product Description

 

Application Areas 
A lot of new technology and new materials are adopted. Such features as LCD display, menu-type instruction operation, curve storage, flameout protector and outside printout are enabled by the application of top grade embedded mono-chip computer circuit.

 

Feature :

Direct concentration read-out

Automatic Calculation of correlation coefficient

Pre-selection of flame sizes

Flameout protection device

Measuring range changing

LPG burning gas

Air compressor including

 

 

 

 

 

 

 

 

 

 

  • RangeK:0~999.9;Na:0~999.9;Li:0~999.9
  • Sensitivity:K:0.2μg/mL;Na:0.2μg/mL
  • Linearity:K:0.2μg/mL;Na:0.7μg/mL
  • Drift:less than 3% within 15s
  • Reproducibility:3%
  • Aspiration rate:<6mL/min
  • Fuel supply:L.P.G.

 

Working conditions

  • Environmental Temperature: 10 oC~35oC
  • Relative humidity:≤85%
  • The instrument should be placed horizontally, avoiding direct sunshine radiation. There is no disturbance of strong electromagnetic field or vibration.
  • Fire extinguishing device should be equipped in the application field and good aeration should be guaranteed. Power supply: AC220V±22V, frequency 50Hz±1Hz,with good grounding.

 

Standard configuration:

  • Main Instrument  1
  • Power Cable  1
  • Chimney  1
  • Clearing rod  2
  • Air compressor  1
  • Printer SP-T 2425SK  1

Optional accessories:

  • Printer SP-T 2425SK
  • Air compressor
  • Lithium filter

 

Technical Parameter

 

 

 

DW-FP640

DW-FP6410

DW-FP6430

DW-FP6431

DW-FP6432

DW-FP6440

DW-FP6450

 

Sensitivity

K

0.156ppm

0.156ppm

0.156ppm

 

 

Na

0.184ppm

0.184ppm

0.184ppm

 

 

Li

------

optional

0.1ppm

------

------

0.1ppm

 

 

Ca

------

------

------

2ppm

------

2ppm

 

 

Ba

------

------

------

------

6ppm

------

6ppm

 

Range

K

0-100ppm

0-100ppm

0-100ppm

 

 

Na

0-160ppm

0-160ppm

0-160ppm

 

 

Li

------

optional

0-100ppm

------

------

0-100ppm

 

 

Ca

------

------

------

0-1000ppm

------

0-1000ppm

 

 

Ba

------

------

------

------

0-3000ppm

------

0-3000ppm

 

Linearity

K

0.195ppm(0.39-3.12)ppm

0.195ppm(0.39-3.12)ppm

0.195ppm(0.39-3.12)ppm

 

 

Na

0.69ppm(1.15-9.2)ppm

0.69ppm(1.15-9.2)ppm

0.69ppm(1.15-9.2)ppm

 

 

Li

------

optional

0.15ppm(0.25-5)ppm

------

------

0.15ppm(0.25-5)ppm

 

 

Ca

------

------

------

3ppm(5-100)ppm

-------

3ppm(5-100)ppm

 

 

Ba

------

------

-------

------

9ppm

------

9ppm

 

Repeatability

 

1 % CV for 20 consecutive samples

1 % CV for 20 consecutive samples

1 % CV for 20 consecutive samples

 

Respond time

 

<8s

<8s

<8s

 

Sample uptake

 

<6L/min

<6L/min

<6L/min

 

Package

 

N.W : 18kg    G.W. 22kg    770 x 560 x 390mm   0.168M3

 

 

 

Detailed Images

 

 

DW-FP640 LCD display Lab equipment of Flame PhotometerDW-FP640 LCD display Lab equipment of Flame PhotometerDW-FP640 LCD display Lab equipment of Flame PhotometerDW-FP640 LCD display Lab equipment of Flame PhotometerDW-FP640 LCD display Lab equipment of Flame PhotometerDW-FP640 LCD display Lab equipment of Flame Photometer 

Send Inquiry to This Supplier

Message
0/5000
Want the best price? Post an RFQ now!
Business Type
Manufacturer, Trading Company
Year Established
2014
Factory Size
1,000-3,000 square meters
Total Employees
101 - 200 People