Mask Medical PFE Test Instrument
1.Product Description
With clean compressed air, the prepared NaCl solution or a similar solution such as DEHS/PAO are sprayed from the aerosol generator. After passing through a classifier (inertia principle), unneeded large particles are collected into the collection bottle, and particles meeting the requirements enter into the test system from the nozzle of aerosol generator.
2. Features
1.The aerosol emitted by the aerosol generator requires two treatments before reaching the tested filter material : 1. Produce aerosol salt particles by mixing with the heated gas (for salt spray). 2. Neutralize the static electricity carried by the aerosol through an electrostatic neutralizer, this treatment is to reduce the impact of particle loss and static electricity on the test results.
2.After the aerosol is processed by the above two treatments, it reaches to the surface of tested filter material at a certain flow rate (flow rate), and then the test procedure is started. For the filter efficiency test, the test time is very short, and it takes only a few seconds to complete the test. For the loading experiment, the experiment is relatively long. The particle counter or laser scattering photometer is used to measure the concentration or mass concentration of the upstream and downstream aerosol, and to calculate the filtration efficiency. At the same time, the pressure drop on both sides of the filter is measured by the differential pressure sensor.
3. Technical Parameters
Flow range | 0-100 L / min, accuracy 2% |
Filtration efficiency | 0.001 ~ 99.999% |
Pressure | 500Pa |
Test area | 100cm² |
Aerosol | NaCl、Oil |
test temperature | 23 ± 5 ℃ |
Relative humidity | 60%±10 |
Count median diameter;CMD | 0.075±0.02um(NaCL)、0.185±0.02um(oily) |
Package dimensions | 1850*1000*830mm |
Package weight | 223kg |
power supply | AC220V 50HZ |
Gas source | Supply pressure above 0.45MPa, flow rate 120L / min |
4. Test Standards:
GB/T 19083-2010 5.4
YY0469-2011 5.6.2 Particle protection
GB/T 32610-2016
GB 2626-2019 6.3
GB 19082-2009 5.7
Send Inquiry to This Supplier
You May Also Like
-
BOJAN Mask PFE Testing Machine With CE CertificateUS$ 20000 - 35000MOQ: 1 Unit
-
Jinan Bojan Face Mask PFE Testing EquipmentUS$ 22000 - 35000MOQ: 1 Unit
-
Jinan Bojan PFE Test InstrumentUS$ 22000 - 35000MOQ: 1 Unit
-
BOJAN Mask PFE TesterUS$ 20000 - 35000MOQ: 1 Unit
-
BOJAN Mask PFE Testing Machine Meet EN149US$ 20000 - 35000MOQ: 1 Unit
-
Particulate Filtration Efficiency (PFE) Tester+Mask Testing MachineUS$ 20000 - 35000MOQ: 1 Unit
-
EN143,EN149 Photometer Method Particle Filtration Efficiency (pfe) TesterUS$ 20000 - 35000MOQ: 1 Set
-
EN149 Photometer Automatic Filter (pfe) TesterUS$ 20000 - 35000MOQ: 1 Set
-
Mask Differential Pressure TesterUS$ 4000 - 4600MOQ: 1 Set
-
ASTMF2100 EN149 Photometer Particle Filtration Efficiency Pfe TesterUS$ 34999 - 35000MOQ: 1 Unit