Home > Instruments & Meters > Optical Lens & Instrument > Optical Lens > GUVx-T1xGS2-3LW10 UV Radiometer 2.0

GUVx-T1xGS2-3LW10 UV Radiometer 2.0

GUVx-T1xGS2-3LW10 UV Radiometer 2.0 photo-1
Negotiable MOQ: 1 Piece (Price negotiable depending on order volume and customization)
Key Specifications
Get Latest Price
Brand Name:
Genicom
Place of Origin:
South Korea
Model Number:
GUVx-T1xGS2-3LW10
Payment & Shipping
Payment Methods:
Port of Shipment:
HK
Delivery Detail:
7 days
Brand Name Genicom
Place of Origin South Korea
Model Number GUVx-T1xGS2-3LW10
GUVx-T1xGS2-3LW10 UV Radiometer 2.0 Description

 Water Environments (10Bar), Display : Absolute Power, Output of Display : Relay, 4-20mA (Option), RS485 (Option)

GUVx-T1xGS2-3LW10 UV Radiometer 2.0 Applications
 UV Purification (Water)
 Water treatment

GUVx-T1xGS2-3LW10 UV Radiometer 2.0 Constitution
 Display, Sensor Probe, 5m Cable


GUVx-T1xGS2-3LW10 UV Radiometer 2.0 Case dimensions

Thread/Length for Mounting
Diameter (mm)
Window (mm)
Wrench Size (mm)
Length (mm)
Weight (g) Case Material
PT3/4" /16mm 30 7 26 62 200 STS 316-L


GUVx-T1xGS2-3LW10 UV Radiometer 2.0 Absolute Maximum Ratings

Parameter
Value
Unit
Remark
Min.
Typ.
Max.
Operating Temperature
-30

85
°C
Sensor Probe
0

50 °C
Display
Power Consumption


5
VA


GUVx-T1xGS2-3LW10 UV Radiometer 2.0 Electro-Optical Characteristics (at 25°C)

Parameter
Value
Unit
Remark
Min.
 Typ.
Max.
Detection Wavelength Range
GUVV-T11GS2-3LW10
230
395 nm

GUVB-T12GS2-3LW10
220
320 nm

GUVC-T11GS2-3LW10
220
280 nm

Detection Power Range
0

100
mW/cm2

* Maximum of detection power : 20mW/cm2 (Option), 100mW/cm2 (Standard), 1000mW/cm2 (Option)