GDVA-404 Multifunctional Transformer CT PT Analyzer 

Tests for Current Transformer :
1. Excitation curve and parameters test
2. Turns ratio test
3. Ratio and phase error test
4. Polarity mark check
5. Winding resistance measurement
6. Secondary loop burden measurement
7. Error line curve test for protection CT
8. Transient CT parameters test
9. CT nameplate guess
10. Saturation hysteresis loop curve measurement
Tests for Potencial Transformer :
1. Turns ratio test
2. Ratio error and phase angle error test
3. Polarity test
4. Secondary burden test
5. Winding resistance test
Applications of CT/ PT Analyzer:
1. CT nameplate check
2. CT parameters check in work burden
3. CT transient parameters analysis
4. CT ratio and phase error calibration
5. PT routine test

Main Specifications
GDVA-404 Multifunctional Transformer CT PT Analyzer
1
Test standards
IEC60044-1, IEC60044-6, GB1208, GB16847, C57.13
2
Power supply
AC220V±10%,50Hz/60Hz±10%
3
Power output
0.1~125V (AC)
4
Current output
0.001~5A(RMS)
5
Power output
300VA
6
Max. knee voltage measurement
45kv
7
Current measurement range
0~10A (automatically change range in 0.1/0.4/2/10A)
8
Current measurement error
9
Voltage measurement range
0~200 V (automatically change range in 1V/10V/70V/200V)
10
Voltage measurement error
11
Turns ratio measurement range
1~30000;

1~2000 error
12
Phase measurement
error:±2min,resolution:0.01min
13
Coil resistance measurement
Range: 0~8kΩ(automatically change range in 2ohm /20ohm /80ohm/800ohm/8kohm)
Error
14
Temperature measurement
-50~100 Celsius degree, error
15
CT Secondary burden
Range 0-160ohm (automatically change range in 2ohm/ 20ohm/80ohm/160ohm)

Error
16
PT Secondary burden
Range 0-80kohm (automatically change range in 800ohm / 8kohm /80kohm)

Error
17
PT ratio measurement
Range: 1~30000, 1~5000 error
18
Saved data groups
>1000groups
19
Work condition
Temperature:-10℃-50℃, moist:≤90%
20
Size
70*50*41CM
21
Weight
40kg


Product Tags: Automatic Ct Pt Analyzer , Ct Pt Testing Equipment , Current Transformer Analyazer

Send Inquiry to This Supplier

Message
0/5000

Want the best price? Post an RFQ now!
Business Type
Manufacturer
Year Established
2009
Annual Export Value
US$2.5 Million - US$5 Million
Port of Shipment
Chongqing, Shenzhen, Guangdong, Hongkong

Recommended Categories

Explore top categories and find suppliers for your specific needs