Bante900 Multifunction Water Quality Meter
Multifunction water quality meter includes 15 optional parameters, the meter is suitable for measuring the
pH, ORP, Ion concentration, Conductivity, TDS, Salinity, Resistivity and Dissolved Oxygen of the liquids.

Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)

Bante900 Multifunction Water Quality Meter - Features:
pH Mode:
• Multifunction water quality meter is equipped with a large backlit LCD display.
• 1 to 5 points push-button calibration with auto-buffer recognition.
• Selectable pH buffer standards (USA/NIST/DIN) or using the custom calibration solutions.
• Automatic electrode slope display helps user decide whether to replace sensor.

Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)

ORP Mode:
• Single point offset calibration allows adjusting the displayed value to a known standard.
• Relative and absolute millivolt measurement modes provide accurate ORP readings.

Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)

Ion Mode:
• 2 to 5 points calibration including the eight concentration calibration points can be selected.
• Automatically recognize ion selective electrodes, does not need to specify type of ion.
• Direct ion concentration readout simplifies the elaborate measurement process.
• Selectable multiple concentration units including the ppm, mg/L and mol/L.

Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)

Conductivity/TDS/Salinity/Resistivity Mode:
• 1 to 5 points push-button calibration with automatic calibration solution recognition.
• Selectable cell constants (0.1/1/10), normalization temperatures (20/25°C), TDS conversion
  factors, seawater and practical salinity measurement modes.

Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)

Dissolved Oxygen Mode:
• 1 or 2 points calibration using air-saturated water and zero oxygen solution.
• Manual salinity and barometric pressure compensation improve the measurement accuracy.

Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)

Other Features:
• Automatic Temperature Compensation provides accurate reading over the entire range.
• Calibration Due Reminder prompts user to calibrate the meter regularly.
• Stability indicator automatically shows current measurement status.
• Auto-Hold function freezes stable measured value for easy viewing and recording.
• Manual temperature calibration corrects temperature deviation.
• Selectable temperature unit (°C or °F) meets different application requirements.
• Automatic electrode diagnosis shows zero point offset and slope of the electrode.
• Help message as a operational guide that helps you quickly begin using the meter.
• System menu allows setting the 15 parameters, including the calibration points, resolutions,
  stability conditions, measurement units, etc.
• Reset feature automatically resumes all parameters back to factory default options.
• Expanded memory stores and recalls up to 500 readings.
• Built-in real-time clock stamps stored data to meet GLP standard.
• Stored data can be transferred into computer by USB communication interface.

Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)

Bante900 Multifunction Water Quality Meter - Specifications:
pH:
• Range: -2.000~20.000pH
• Accuracy: ±0.002pH
• Resolution: 0.1, 0.01, 0.001pH
• Calibration Points: 1 to 5 points, USA, NIST, DIN Standards
• Temperature Compensation: 0~100°C, 32~212°F, Manual or Automatic

mV:
• Range: -1999.9~1999.9mV
• Accuracy: ±0.2mV
• Resolution: 0.1, 1mV
• Calibration Points: 1 point (Only for Relative mV mode)
• Calibration Range: ±200mV

Ion:
• Range: 0.001~19999ppm, mg/L, mol/L (Depending on range of ISE)
• Accuracy: ±0.5% F.S (Monovalent), ±1% F.S (Divalent)
• Resolution: 0.001, 0.01, 0.1, 1
• Calibration Points: 2 to 5 points
• Calibration Solutions: 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mol/L, mg/L

Conductivity:
• Range: 0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
• Accuracy: ±0.5% F.S
• Calibration Points: 1 to 5 points
• Calibration Solutions: 10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm
• Temperature Compensation: 0~100°C, 32~212°F, Manual or Automatic
• Temperature Coefficient: 0.0~10.0%/°C
• Compensation Modes: Linear or Pure Water
• Cell Constant: K=0.1, 1, 10
• Normalization Temperature: 20 or 25°C

TDS:
• Range: 0~10ppt (Max. 20ppt, depending on factor setting)
• Accuracy: ±1% F.S
• TDS Factor: 0.1~1.0 (Default 0.5)

Salinity:
• Range: 0~10ppt (Max. 80ppt)
• Accuracy: ±1% F.S
• Measurement Modes: Seawater or Practical Salinity

Resistivity:
• Range: 0~100MΩ
• Accuracy: ±1% F.S
• Resolution: 0.01, 0.1, 1

Dissolved Oxygen:
• Range: 0.00~20.00mg/L (or ppm)
• Accuracy: ±0.2mg/L
• Resolution: 0.01mg/L
• Calibration Points: 1 or 2 points
• Temperature Compensation: 0~50°C, 32~122°F
• Pressure Correction: 60.0~112.5kPa, 450~850mmHg
• Salinity Correction: 0~50g/L

% Saturation of Oxygen:
• Range: 0.0~200.0%
• Accuracy: ±2.0%
• Resolution: 0.1%

Temperature:
• Range: 0~105°C, 32~221°F
• Accuracy: ±0.5°C, ±0.9°F
• Resolution: 0.1°C
• Calibration Points: 1 point
• Calibration Range: Measured value ±10°C

Others:
• Hold Function: Manual or Automatic
• Stability Conditions: Low or High
• Calibration Due: 0 to 31 days
• Power Off: Manual or Automatic (10, 20, 30 minutes)
• Reset Function: Yes
• Memory: Stores up to 500 data sets
• Output: USB Communication Interface
• Power Requirements: DC5V, using AC adapters, 220VAC/50Hz
• Dimensions: 210(L)×188(W)×60(H)mm
• Weight: 1.5Kg

Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)

Bante900 Multifunction Water Quality Meter - Accessories:
• pH Electrode
• Conductivity Electrode
• Dissolved Oxygen Probe
• Temperature Probe
• pH Buffer Solutions (pH4.01, 7.00, 10.01)
• Calibration Solutions (84µS/cm, 1413µS/cm, 12.88mS/cm)
• DO Filling Solution
• DO Cap Membrane
• USB Cable
• DC5V Power Adapters


Bante900 Multifunction Water Quality Meter (Containing the 9 measurement modes)
Bante Instruments - Professional Manufacturer for Multifunction Water Quality Meter

Send Inquiry to This Supplier

Message
0/5000

Want the best price? Post an RFQ now!
Business Type
Manufacturer
Year Established
2010
Factory Size
1,000-3,000 square meters
Total Employees
11 - 50 People

Recommended Categories

Explore top categories and find suppliers for your specific needs