ECG Signal Simulator
ECG Waveform Simulator
Functional Features
l 10 universal connection head, can connect any ECG products.
l 12 leads synchronous ECG signal output, output different 12-lead ECG waveform.
l Built-in high performance CPU, can simulate the output of standard heart rate of 60bpm ECG waveform.
l Built-in large capacity lithium battery, external special charger.
l The system has automatic dormancy function, battery overrush, overdischarge protection function.
l Digital tube display waveform code, frequency, amplitude and other parameters content.
l Key operation, change the waveform type, waveform frequency, amplitude and other parameters.
Type of waveform
1. Collect normal ECG waveform of human body
2. Standard heart rate waveform (sharp Angle wave), used to detect the heart rate range
3. Irregular heartbeat waveform
4. Square wave detection ECG low frequency characteristics and scanning speed
5. Sine wave detection of high frequency ECG performance
6. Impedance breathing of respiratory waveform detection monitor
7.1mV calibration signal detection range
8. Set waveform amplitude parameters
9. Change the base width of the triangular waveform
10. QRS waveform of the pre-pacing signal
11. Standard QRS waveform
12. Pacing signals
13. Sine wave of superimposed triangular waveform
Model configuration Sheet
Model number (Model) |
SKX-2000A |
SKX-2000C |
SKX-2000D |
SKX-2000G |
Waveform Configuration (Configuration) |
60bpm |
1-2 |
1-7 |
1-13 |
Waveform parameter introductionAmplitude error ≤5%, frequency error ≤2%
1, Normal ECG waveform:
★ Heart rate setting range: 10bpm-200bpm (initial value: 60 bpm)
★ The signal amplitude is fixed
2, Positive and negative triangular waveform:
★ Frequency range: 10bpm-400bpm; (Initial value: 75 bpm)
★ Amplitude range: 0.1mV-4mV (10:0.1 mV, 400:4mV)
★ 2 modes, mode 1: positive waveform, mode 2: negative waveform; Select by key combination
3, Simulate irregular heart rate waveform:
★ Frequency range: 10bpm-250bpm; (Initial value: 80 bpm)
★ Amplitude range: 0.1mV-4mV (10:0.1 mV, 400:4mV)
★ 2 modes, mode 1: large waveform below, mode 2: large waveform above; Select by key combination
4, Square Wave:
★ Frequency range: 0.1Hz-10Hz (10:1Hz 100:10Hz) (initial value: 1Hz)
★ Amplitude range: 0.1mV-4mV (10:0.1 mV,400:4mV)
5, Sinusoidal waveform:
★ Frequency range: 1Hz-100Hz (initial value: 10 Hz)
★ Amplitude range: 0.1mV-4mV (10:0.1 mV,400:4mV)
★ The DC component can be superimposed on the waveform, the range is 0.1mV-3mV (10:0.1mV, 300:3mV)
★ Through the combination of keys, select the parameters that need to be changed, mode 1: frequency; Mode 2: DC component of superposition (10:0.1 mV, 400:4mV)
6, Analog breathing waveform (may have no signal for some models) :
★ Frequency range: 1-100 times/min (initial value: 15 bpm)
★ Please note that the respiratory lead is RA-LL and the baseline impedance is 1K
7, Calibration waveform 1MV:
★ Frequency range: 30-60bpm
★ Pulse width: 100ms
8, Signal amplitude setting:
★ Amplitude range: 0.1mV-4mV (10:0.1 mV, 400:4mV) (initial value: 100)
★ Please note that this amplitude setting will affect all waveforms
9, Ecg waveforms with a pacemaker signal:
★ 2 working modes: Mode 1: forward ECG signal with pulse; Mode 2: negative ECG signal with pulse pulse; Selection by key combination
★ Pulse waveform width: 1ms-30ms (initial value: 30ms)
★ Frequency range: 20-250bpm (initial value: 80bpm)
★ Pulse waveform width adjustment, affecting the pulse waveform in two modes
★ The amplitude of the pulse waveform is affected by the signal amplitude setting of 8
A, Triangle waveform with adjustable width:
★ Waveform amplitude: 0.1mV-4mV (10:0.1mV, 400:4mV)
★ Frequency range: fixed 60 times /s
★ Width adjustment range: 2ms-400ms (initial value: 100ms)
b, analog ECG waveform:
★ This waveform has 4 working modes, mode 1: can change the amplitude of T wave; Mode two: set the waveform frequency; Mode 3: set the amplitude of QRS waveform; Mode four: set the width of the QRS waveform, and convert between the working modes by the combination of keys
★ T wave amplitude setting range: 0.01mV-2mV (initial value: 10:0.1 mv)
Waveform frequency setting range: 20bpm-300bpm (initial value: 75bpm)
★ QRS waveform amplitude range: 0.1mV-2mV (initial value: 1mV)
★ QRS waveform width range: 10ms-150ms (initial value: 80ms)
C, starting signal (pulse waveform) :
★There are 2 kinds of this waveform, respectively, the pulse waveform in the positive and negative direction, and the edge change is 300us
★ The unit is 0.1 milliseconds
★ The amplitude size needs to be set in the signal amplitude setting (8)
d, superimposed 0.5Hz triangular wave sine waveform:
★ The amplitude of the triangular wave can be set
Send Inquiry to This Supplier
You May Also Like
-
Patient SimulatorNegotiableMOQ: 1 Set
-
ECG Simulator for ECG Performance TestingNegotiableMOQ: 1 Set
-
ECG Database Player Designed for System Validation and Software Algorithm EvaluationNegotiableMOQ: 1 Set
-
CMRR TesterNegotiableMOQ: 1 Set
-
EEG Performance Tester Designed for Regulation TestingNegotiableMOQ: 1 Set
-
EEG Performance TesterNegotiableMOQ: 1 Set
-
Health Wareable TestingNegotiableMOQ: 1 Set
-
Wearable Devices ECG TesterNegotiableMOQ: 1 Set
-
NIBP Simulator & AnalyzerNegotiableMOQ: 1 Set
-
AED/Defibrillator TesterNegotiableMOQ: 1 Set