HPHT Consistometer With IPRO Control System
| HPHT Consistometer with iPRO Control System |
|

The High-Pressure, High-Temperature (HPHT) Consistometer measures cement slurry viscosity or consistency under elevated pressure and temperature conditions. Its primary function is to determine the maximum available pumping time of a cement slurry before the slurry reaches an un-pumpable consistency before setting. While designed for cement slurries, the effects of pressure, time and temperature can also be observed for other fluids, emulsions, dispersions, or slurries under static or dynamic conditions.
Operation
The FANN Model 275 HPHT Consistometer exposes a cement slurry sample to a controlled set of parameters of temperature, agitation and pressure that simulate the down-hole conditions. During a test, these parameters are carefully monitored and precisely controlled by the iPRO Control System Software.
A flat panel, High-Resolution, LCD touch-screen provides the interface between the user and the software. The LCD screen provides real-time viewing of temperature, pressure, and consistency data in graphic and text formats. This data is automatically recorded to a database for future analysis to help predict the cement slurry's down-hole performance.
Hardware
The Model 275 HPHT Consistometer is composed of several subsystems controlled and coordinated by System Control software. The iPRO Control system utilizes a Data Acquisition Board for a Real-time data control and acquisition.

iPRO Consistometer Control Software is the next generation of complete and powerful system control software with a simple and friendly, flexible user interface.
-
Touch Screen with Friendly and Flexible User Interface
- Unlimited storage for test data
- Remote Test control & monitor via LAN, Intranet, and Internet
- Email test results from within software
Basic Features
- Temperature up to 400°F (204° C)
- Pressure to 30,000 psig (207 MPa)
- Slurry cup rotation - 150 RPM per API specifications
- Metal-to-Metal pressure chamber closure - ease of operation, less maintenance
- Pressure chamber cooling coils - fast cool-down, quick cleanup
- Ceramic-insulated electrical connections through pressure chamber
Safety Features
- Rupture disc for accidental chamber overpressure
- Automatic power shutdown if test temperature exceeds safe perating limits or if a break occurs in the temperature controlling thermocouple
Ordering Information
Part No. 209599 - HPHT Consistometer Model 275
Requirements
| Electrical Power Supply | 230 Volts |
| Current | 30A |
| Frequency | 60 Hz or 50 Hz (specify) |
| Cooling Water Supply | 30 psig min. (0.207 mPa) |
| Compressed Air Supply | 90 psig min. (0.621 mPa) |
| Drain for Cooling Water |
Send Inquiry to This Supplier
You May Also Like
-
Resistivity MeterNegotiableMOQ: 1 Piece
-
RheometerNegotiableMOQ: 1 Piece
-
Shearing Mixer / DispersatorNegotiableMOQ: 1 Piece
-
Static Fluid Loss TestNegotiableMOQ: 1 Piece
-
TRU-WATE Pressurized Fluid Density Scale,tru-wate Pressurized Mud BalanceNegotiableMOQ: 1 Piece
-
Aging Oven,aging Roller OvenNegotiableMOQ: 1 Piece
-
Hot Roller Oven for Drilling Fluids Testing EquipmentsUS$ 100 - 500MOQ: 1 Piece
-
HTHP Filter Press for Drilling Fluids Testing EquipmentsUS$ 100 - 500MOQ: 1 Piece
-
Aging Roller Oven,aging Cells,drilling Fluids Testing Equipments,cement Testing EquipmentsUS$ 100 - 500MOQ: 1 Piece
-
Retort Kit 50ml for Drilling Fluids Testing EquipmentsUS$ 100 - 500MOQ: 1 Piece