HK-1003C Automatic Distillation Apparatus for Petroleum Products
Overview
HK-1003C automatic distillation apparatus for petroleum products is a new designed distillation tester. It conforms to standard ASTM D86 standard Test Method for Distillation of Petroleum Products at Atmospheric Pressure, which is suitable for testing distillation of light and middle distillates, automotive spark-ignition engine fuels with or without oxygenates, aviation gasoline, aviation turbine fuels, diesel fuels, biodiesel blends up to 20 %, marine fuels, special petroleum spirits, naphthas, white spirits, kerosines, and Grades 1 and 2 burner fuels.
Features
- The apparatus can automatically measure and control all experimental process by the single chip microcomputer, 10-inch touching LCD screen displays all experimental process and show the temperature data, volume data and related curves.
- Our level tracking system consists of American brand Haydon High subdivision step linear motor, KEYENCE (JAPAN) ball type circular positioning screw and Imported linear laser tracker.
- We adopt mechanical refrigeration for the cool tube and cylinder tube. We use brand Danfoss (Secop) compressor as our Inner loop cooling medium system, user can detect and add the cooling water every 2 years.
- The apparatus can automatically control the distillation heating circuit, and makes the sample flow rate control from the initial distillation point to 95% within 4 ~ 5 ml per minute.
- The apparatus can provide initial distillation point temperature and final distillation point temperature value, also can provide temperature and flow rate at percent point.
- The apparatus can detect local atmospheric pressure automatically, and correct it to the temperature under standard atmospheric pressure
- User can choose to crease the test according to the steam temperature
- The apparatus can store and detect data directly; also the data can be printed.
- The baffle adopts quartz, which can gives high anti-oxidation thermal efficiency up to 1000℃
Technical parameters
- Rated voltage: AC220V±10% 50Hz
- Power: 1200w
- Temperature range: Environment temperature~400°C
- Temperature control Precision:±0.1°C
- Temperature sensor:RTD Pt100
- Start heating to Initial distillation point time: 5~15min
- Distillation rate: 4~5min
- Residues in flask is 5 ml; less than 5 min
- Optical fiber sensors detect initial distillation point temperature
- Recycling volume measurement range: 0~100ml
- Recycling volume measurement precision:±0.1ml
- Ambient requirements: temperature:15°C~40°C Humidity≤85%
Send Inquiry to This Supplier
You May Also Like
-
OZ7901 Membrane-Based Dissolved Ozone Electrode for Water TreatmentNegotiableMOQ: 100 Pieces
-
Water Quality Analysis ElectrolyzerUS$ 3 - 3.5MOQ: 50 Pieces
-
LIYI RoHS Testing X-Ray Fluorescence Spectrometer Energy Dispersive XRF Analyzer PriceUS$ 14,580 - 21,450MOQ: 1 Set
-
Fish Aquaculture Oxygen EquipmentUS$ 999 - 1,111MOQ: 1 Piece
-
Winner3005 Dry Particle Micron Measuring Instruments for Abrasive Powder TestNegotiableMOQ: 1 Set
-
AMT-3 Mineral Detect InstrumentUS$ 100 - 300MOQ: 1 Set
-
Laser Measurement SensorNegotiableMOQ: 1 Set
-
SELON THERMAL CONDUCTIVITIY METER, DIGITAL PH METER,PH METER DIGITALUS$ 587 - 587MOQ: 1 Piece
-
Whiteness TesterNegotiableMOQ: 1 Unit
-
100% Factory Selling of Garden 2 in 1 Soil ph Test MeterNegotiableMOQ: 500 Pieces