Home > Instruments & Meters > Test Instrument > Test Machine > MS7207 High Accuracy Digital Loop Calibrator Process Calibrator 0.025% Accurancy

MS7207 High Accuracy Digital Loop Calibrator Process Calibrator 0.025% Accurancy

MS7207 High Accuracy Digital Loop Calibrator Process Calibrator 0.025% Accurancy photo-1
US$ 130 - 180 MOQ: 1 Piece
Key Specifications
Get Latest Price
Place of Origin:
Guangdong, China
Brand Name:
Graigar
Model Number:
MS7207
Payment & Shipping
Payment Methods:
Port of Shipment:
Shenzhen
Delivery Detail:
Delivery time depends on order quantity.
Place of Origin Guangdong, China
Brand Name Graigar
Model Number MS7207

Specifications

MS7207 High Accuracy Digital Loop calibrator process calibrator 0.025% accurancy

 

Product Description

 

 

MS7207 High Accuracy Digital Loop calibrator process calibrator 0.025% accurancy

 

MS7207 High Accuracy Digital Loop calibrator process calibrator 0.025% accurancy

 

  Loop  Calibrator   MS7207

Features
The fast, one-handed tool for loop checks. Just click to calibrate.
This feature-rich solution combines the best features push-button and knob-base loop calibrator and offers:
·         Large display and simple, quick-click interface for easy one-handed operation.
·         Simultaneous mA and % readout for quick, easy interpretation of readings.
·         mA accuracy of 0.025%, superior to other loop calibrators in this price range.
·         1 µA resolution for mA source, simulate and measure.
·         Push button 25% steps for fast, easy linearity checks.
·         0-100% “span check” for fast confirmation of zero and span.
·         Selectable slow ramp, fast ramp, and step ramp to provide smooth outputs for valve slewing and loop functional tests.
·         24 volt internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM.
·         Measures V dc to 28 V.
·         Single 9V battery that is easily changed without braking calibration seal
·         0 - 20 mA or 4 - 20 mA default start up modes.
·         Innovative output adjustment dial with 1 µA and 100 resolution.
·         HART ™ mode connects 250 Ω resistor in series with 24 V loop for compatibility with HART communicators.

technical_parameters

 

Specifications

Current Measurement Range: 0 to 24 mA
Resolution: 0.001 mA
Accuracy: 0.025% Rdg + 2 LSD
Current Sourcing Range: 0 to 20 mA or 4 to 20 mA
Accuracy: 0.025% Rdg + 2 LSD
Drive Capability: 1200 Ω; 950 Ω at 20 mA in HART mode
Current Simulation Range: 0 to 20 mA or 4 to 20 mA
Accuracy: 0.025% Rdg + 2 LSD
Ext. loop voltage requirement:  12-30 V
Loop Power  While measuring mA, 24 V
Voltage measurement Range: 0 to 28 V DC
Resolution: 1 mV
Accuracy: 0.025% Rdg + 2 LSD

 

 MS7207 High Accuracy Digital Loop calibrator process calibrator 0.025% accurancy

 

Our Services

 

 

 

MS7207 High Accuracy Digital Loop calibrator process calibrator 0.025% accurancy

 

 

 

 

 

 

 

 

 

 

 

 

Send Inquiry to This Supplier

Message
0/5000

Want the best price? Post an RFQ now!
Business Type
Manufacturer, Trading Company
Year Established
2012
Factory Size
1,000-3,000 square meters
Total Employees
11 - 50 People

Recommended Categories

Explore top categories and find suppliers for your specific needs