Home > Instruments & Meters > Test Instrument > Test Machine > Flash Fire Point Tester

Flash Fire Point Tester

Flash Fire Point Tester photo-1
US$ 380 - 420 MOQ: 1 Set
Key Specifications
Get Latest Price
Place of Origin:
Shanghai, China
Brand Name:
SHTG
Model Number:
SYD-3536
Payment & Shipping
Payment Methods:
Port of Shipment:
Shanghai port
Delivery Detail:
Delivery time depends on order quantity.
Place of Origin Shanghai, China
Brand Name SHTG
Model Number SYD-3536
Power Electronic
Usage Other, Cleveland Open Cup Flash Point Tester

Specifications

flash fire point tester
1.AC 220V, 50 Hz
2.Not more than 650 W

flash fire point tester

 

Purpose

The instrument is designed and made as per the National Standard of People’ Republic of China GB/T3536-2008 “Test Methods for Flash and Fire Point of Petroleum Products (Cleveland Open Cup Methods)”, and the Company Standard of Shanghai Q/YXYY15 “SYD-3536 Cleveland Open Cup Flash Point Tester”. It is suitable to determine flash point and fire point of petroleum products having a flash point higher than 79 °C, excepting fuel oils.
 
specification 
1. Power supply: AC 220V±10%, 50 Hz;
2. Heating device: It is a quartz tube heater, so there is no flame appears. Explosion protective; its power can be continuously adjusted in the range of 0∼600W;
3. Flame application device: automatically scan;
4. Thermometer: -6~400°C; Each scale division is 2 °C; as per GB/T514;
5. Ignition device:
(1) Ignition source: coal gas (or any other combustible gas);
(2) Spout aperture: 0.6∼0.8 mm;
6. Ambient temperature: -10°C∼50°C;
7. Relative humidity: ≤85%;
8. Total power consumption: not more than 650 W;
 
flash fire point tester

 

Send Inquiry to This Supplier

Message
0/5000

Want the best price? Post an RFQ now!