Specifications
1.real-time detection carbon monoxide2. 0~10VDC/4~20mA output
3. Modbus RS485 communication
4.CE-approval

Real time monitoring carbon monoxide transmitter
Sensors | ||
Gas Sensor | Electrochemical carbon monoxide sensor | |
Sensor lifetime | Up to 5 years, replaceable | |
Warm up time | 1hour (first time) | |
Response Time | Within 60 seconds | |
Signal Update | 1s | |
CO Measuring Range | 0~500ppm(default)/0~100ppm/0~1000ppm selectable | |
Accuracy | <1ppm+5%reading | |
Stability | ±5% (over 900 days) | |
Temperature & Humidity Sensor (optional) | Temperature | Relative Humidity |
Sensing element: | Band-gap-senor | Capacitive humidity sensor |
Measuring range | -20°C~60°C | 0 -100%RH |
Accuracy | ±0.5°C (20~40°C) | ±4.0%RH (25°C, 15%-85%RH) |
Display resolution | 0.1°C | 0.1%RH |
Stability | ±0.1°Cper year | ±1%RH per year |
Outputs | ||
LCD Display (optional) | Display COmeasurement or CO+ temperature& humidity measurements | |
Analog Output | 1X0~10VDC or 4~20mA linear output | |
Analog Output Resolution | 16Bit | |
Relay dry contact Output | Up to two dry-contact outputs Max, switching current 3A (230VAC/30VDC), resistance Load | |
Modbus Interface | Modbus RS-485 with 19200bps(default), other rates selectable in the order. 15KV antistatic protection | |
Electrical and General Items | ||
Power Supply | 24VAC/VDC | |
Power Consumption | 2.8W | |
Wiring Standard | Wire section area<1.5mm2 | |
Working Condition | -10°C~60°C(14~140°F); 5~99%RH, non condensing | |
Storage Conditions | -10~60°C(14~140°F)/ 5~99%RH, non condensing | |
Net Weight | 260g | |
Manufacturing Process | ISO 9001 Certified | |
Housing and IP class | PC/ABS fireproof plastic material, protection class: IP30 | |
Compliance | CE-EMC Approval | |
1.Design for real-time detection indoor carbon monoxide.
2.High accuracy humidity and temperature detection optional
3.LCD display carbon monoxide and optional temperature&RH measurement
4.Smart buttons for easy operation
5.Excellent electrochemical CO sensor with up to 5 years life time
6.Replaceable CO sensor
7.Provide 1X analog linear output (0~10VDC/4~20mA selectable) for the measurement
8.Providing up to two dry contact outputs with presetting the setpoint
9.Modbus RS485 communication interface optional
10.Provide two easy methods for calibration and setup alarm points by end users: infrared remote controller or a software via RS485
11.24VAC/VDC power supply
12.CE-Approval
8.24VAC/VDC power supply
Send Inquiry to This Supplier
You May Also Like
-
Qualified HST CS Series Iron Ore Carbon Sulfur AnalyzerNegotiableMOQ: 1
-
Supply of Electric Carbon Products Carbon and Sulfur Analyzer, Mingrui mr - Cs995 Type, High - Frequency Infrared Carbon and Sulfur AnalyzerUS$ 6123.08 - 6584.62MOQ: 1 Unit
-
Honeywell EZtrend® GR Advanced Graphic Recorder AgentNegotiableMOQ: 1 Unit
-
Lora Wireless CO2 SensorUS$ 165 - 180MOQ: 1 Piece
-
TOC Offline AnalyzerNegotiableMOQ: 1 Set
-
DI1500 Water Detection Industrial Laboratory TOC Analyzer Price Total Organic Analyzer MeterUS$ 10,000 - 15,000MOQ: 1 Piece
-
Online Technique Support Toc Analyzer With Good PriceUS$ 1,000 - 15,000MOQ: 1 Piece
-
CONTACT BLOCK FOR TEMPERATURE AND OXYGEN SENSORS-4(CORE)OXY&2(CORE)TEMP. USED IN MEASURING OXYGEN&TEMPERATURE ACTIVITY OF MOLTEN METALS IN TYPES B,R,S&IN MARK VII MODELS.NegotiableMOQ: 1 Piece
-
Non Destructive Testing Glass Miniature X Ray Tube With Tungsten FilamentNegotiableMOQ: 1 Piece
-
20X 40X Jewelry Gem Stereo Optical Microscope Parallel Microscopes A22.1207US$ $1 - $1MOQ: 1 Piece