Optical Emission Spectrometer HWLD-4C
Optical Emission Spectrometer HWLD-4C
Features
1 Various metal bases can be analyzed: ferrous metals of Fe, Co, Ni and Ti base, and non-ferrous metals of Cu, Al, Pb,Mg, Zn and Sn base
2 Adopting the exciting source with high reproducibility and stability. The exciting frequency can be changed in the range of 150-600Hz to match the material to be analyzed and get the best analytical results
3 Adopting highly integrated data collecting and controlling system with high automation
4 High-speed analysis with high antijamming ability, good reproducibility and stability
5 Reasonable design with more integrated construction
6 Windows based software with easy and simple operation
7 In-built data base allowing long-distance data transfer
8 Quality control diagram can be established automatically for easy quality control and monitoring
9 High vibration resistance ability. It's not necessary to make vibration resistance base
10 Partial thermostat system for fewer requirements to the environment
Specifications
Spark Source
Power supply: AC 220v ±10% 50 Hz
Input power: 1.0 kVA
Spectroscope
Spectral range: 175∼450 nm
Grating: 2400L/mm concave grating in radius of 0.75m; Blazed at 300nm (1st order); Ruled area: 30×50mm2; Reciprocal linear dispersion: 0.55nm/mm
Width of entrance slit: 20 µm
Width of exit slit: 50 µm, 75 μm
Maximum number of channel: 36
Constant temperature: 30±0.1 °C
Measurement and control system
Measuring mode: sectional integration
Measuring reproducibility: RSD ≤ 0.2%
PMT high voltage power supplies: Voltage -1000v
Stability: better than0.5% (8 hours)
PMT high voltage adjustment: controlled by PC
Data Processing System
Industrial controlled built in PC, LCD display. Software: under Windows Operating system
Dimensions
Main Unit: 1500 (l) ×1100 (w) × 1200 (h) mm3
Area required: 15-20 square meters
Weight: about 450 kg
Laboratory requirements
Environment temperature:20 ±5°Crelative humidity:<70%
Purity of Argon gas: 99.999%
Power supply: AC 380V 10%, 50Hz, 3KW, 3-phase, 4-line
Ground lead: An independent ground lead should be used for the spectrometer itself. Grounding wire resistance< 2ohm
Send Inquiry to This Supplier
You May Also Like
-
Endress+Hauser E+H Compact Conductivity Device Smartec CLD18-BGGENegotiableMOQ: 1 Piece
-
100kn Universal Cable Tensile Testing Machine Benchtop 100kn Biaxial Briquettes V Belt Universal Tensile Testing MachineUS$ 3000 - 10000MOQ: 1 Set
-
Near Infrared Food Ingredient AnalyzerUS$ 11000 - 15000MOQ: 1 Unit
-
HST Spring Dyno Friction Leakage Fatigue Testing MachineUS$ 5000 - 10000MOQ: 1 Set
-
Hydraulic Plastic Flange Connections Cover Plugs SHF SeriesUS$ 0.22 - 0.28MOQ: 2500 Pieces
-
Gates 550C 550U U550 U-550 U508 U-508 Japan Unitta U-550 Belt Tension Meter New U-508 Belt Tension Meter Upgraded Version U550US$ 1 - 900MOQ: 1 Piece
-
Analytical Balanace.scale ,weichengya Laboratory SupplierUS$ 10 - 35MOQ: 10 Sets
-
Smart MD-3051 Online Density Meter With 4-20mA OutputUS$ 2,380 - 2,500MOQ: 1 Unit
-
5W-300W Sun Simulator Watt Testing 30% Off / Xenon Lamp Sun Simulator Free Sent / Sun Simulator TestingUS$ 7,500 - 10,000MOQ: 1 Set
-
New Design Measuring Meter With Best PerforanceUS$ 1,000 - 1,200MOQ: 1 Piece