Steam Sterilizer

Steam Sterilizer photo-1
US$ 20000.00 - 20000.00 MOQ: 1 Unit
Key Specifications
Get Latest Price
Place of Origin:
Shandong, China
Brand Name:
Shinva
Model Number:
G/D series
Payment & Shipping
Payment Methods:
Port of Shipment:
Qingdao
Delivery Detail:
Delivery time depends on order quantity.
Place of Origin Shandong, China
Brand Name Shinva
Model Number G/D series
Application Other, all
Type Other, pressure vessel
After-sales Service Provided Engineers available to service machinery overseas
Product Description

 

Pulse Vacuum Sterilizer is of steam sterilizer of high performance, which can choose the sterilizing cycle with the automatic control. As well as the advanced sterilizing device that in accord with GMP regulation. It is widely used for the sterilization of biological products, glass vessels, aseptic clothes, medical instruments and dressing, which require good sterilization in areas of biology engineering, medical hygiene, animal laboratory and pharmaceutical factory.

 

 

l  GB 150.1~GB 150.4-2011  Steel Pressure Vessel

 

l  GB/T 191-2008  Pictorial Markings for Handling of Packages

 

l  GB 4793.1-2007  Measurement, Control and Lab Safety Requirements of Electrical Equipment, Part 1: General Requirement

 

l  GB 9969.1-1998  General Principles for Preparation of Instructions for Use of Industrial Products

 

l  GB/T 14710-2009  The Environmental Requirements and Test Methods for Medical Electrical Equipment

 

l  GB 18278-2000  Sterilization of Health Care Products--Requirements for Validation and Routine Control--Industrial Moist Heat Sterilization

 

l  JB/T 4711-2003  Coating, Shipping and Packing of Pressure Vessel

 

l  YY 0076-1992   Metal Parts Classification Technology Conditions of Coating

 

l  YY 0085.1-1992 Pulse Vacuum Pressure Steam Sterilizer

 

l  SFDA  2010 version GMP

 

l  Chinese Pharmacopoeia 2010

 

l  Supervision Regulation on Safety Technology for Fixed Type Pressure Vessel

 

l  ISPEVolume 4, Water and Steam System, Edition 1, Jan. 2001

 

1      Main performance parameters

 

1.1      Exposure temperature: 105~134;

 

1.2      Chamber working pressure: -0.1~0.23 MPa;

 

1.3      Jacket working pressure: 0~0.23 MPa;

 

1.4      Working noise: less than 80dB;

 

1.5      Surface temperature: not exceed environment temperature 25

 