Home > Health & Medicine > Veterinary > Veterinary Instruments > Professional Veterinary Patient Monitor, Veterinary Equipment

Professional Veterinary Patient Monitor, Veterinary Equipment

Professional Veterinary Patient Monitor, Veterinary Equipment photo-1
US$ 500 - 600 MOQ: 1 Unit
Key Specifications
Get Latest Price
Place of Origin:
Guangdong, China
Brand Name:
Jumper
Model Number:
JPD-800B
Payment & Shipping
Payment Methods:
Port of Shipment:
Shenzhen
Delivery Detail:
Delivery time depends on order quantity.
Place of Origin Guangdong, China
Brand Name Jumper
Model Number JPD-800B
Properties Other, Diagnosis & Injection

Specifications

Veterinary equipment.
1.12.1 inch color touch screen.
2.CE approved.
3.ECG , SPO2 , PR , NIBP , RR , Temp..

Veterinary Patient Monitor(JPD-800B) 12.1 inch. 

 

Feature:

12" color TFT display, up to 12 waveforms

 

One level menu, easy to use

 

Parameters including ECG/RESP, SpO2 , NIBP, TEMP, Pulse Rate and IBP(optional)

 

Maximum 720 hour graphic and tabular trends of all parameters Built-in recorder and battery

 

Suitable for adult, pediatric and neonatal patient

 

Arrhythmia analysis, ST calculation, Pace analysis

 

Anti-ESU, anti- defibrillator

 

Can connected to hospital controling system.

 

Optinal: recorder, dual IBP, CO2, IoC

 

professional veterinary patient monitor, veterinary equipment

Specification:

ECG

Lead mode: 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)

Lead selection: I, II, III, avR, avL, avF, V,

Waveform :2 ch

Lead mode: 3 Leads ( R, L, F or RA, LA, LL)

Lead selection: I, II, III,

Waveform: 1 ch

Gain: X2.5mm/mV, X5.0mm/mV, X10mm/mV, X20mm/mV, auto

Resolution: 1 bpm

Sensitivity: > 200 (uV P-P)

Differential Input Impedance: > 5 MΩ

CMRR

Monitor: > 105 dB

Operation: > 105 dB

Diagnosis: > 85 dB

Electrode offset potential :300mV

Leakage Current: < 10 uA

Baseline Recovery :< 3 S After Defi.

ECG Signal Range: 8 m V ( Vp-p )

Bandwidth

Surgery: 1 ~ 15 Hz

Monitor: 0.5 ~ 35 Hz

Diagnostic: 0.05 ~ 100 Hz

Calibration Signal 1 (mV p-p), Accuracy : ± 5%

Measure and Alarm: -2.0 ~ +2.0 mV

ARR Detecting

Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, 

RON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC

Alarm: Available

 

RESPARATION

Method: Impedance between R-F(RA-LL)

Differential Input Impedance: >2.5 MΩ

Measuring Impedance Range: 0.3~5.0Ω

Base line Impedance Range: 0 ~ 2.5 KΩ

Bandwidth: 0.3 ~ 2.5 Hz

Resp. Rate

Measuring and Alarm Range

Adult: 0 ~ 120 rpm

Neo/Ped: 0 ~ 150 rpm

Resolution: 1 rpm

Accuracy:  2 rpm

Apean Alarm: 10 ~ 40 s

 

NIBP

Method: Oscillometric

Mode: Manual, Auto, STAT

Measuring Interval in AUTO Mode:1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)

Measuring Period in STATMode: 5 Min

Pulse Rate Range: 40 ~ 240 bpm

Alarm Type: SYS, DIA, MEAN

Measuring and alarm range

Adult Mode:

DIA:10 ~ 215 mmHg

MEAN: 20 ~ 235 mmHg

Pediatric Mode

SYS: 40 ~ 200 mmHg

DIA: 10 ~ 150 mmHg

MEAN: 20 ~ 165 mmHg

Neonatal Mode

SYS: 40 ~ 135 mmHg

DIA: 10 ~ 100 mmHg

MEAN: 20 ~ 110 mmHg

Resolution Pressure :1mmHg

Accuracy Pressure Maximum Mean error: ±5mmHg

Maximum Standard deviation: ± 8mmHg

Overpressure Protection

Adult Mode:297 ± 3 mmHg

Pediatric Mode:240 ± 3 mmHg

Neonatal Mode:147 ± mmHg

 

SpO2

Measuring Range: 0 ~ 100 %Alarm Range: 0 ~ 100 %

Resolution: 1 %

Accuracy:70% ~ 100%  ± 2 %    0% ~ 69% unspecified

Actualization interval: about 1 Sec.

Alarm Delay: 10 Sec.

 

Pulse Rate

Measuring and Alarm Range:0~254bpm

Resolution:1bpm

Accuracy :± 2bpm

 

TEMPERATURE

Channel 2

Measuring and Alarm Range: 0 ~ 50 C

Resolution: 0.1 C

Accuracy :  0.1 C

Actualization interval:about 1 Sec.

Average Time Constant: < 10 Sec.

 

 

IBP(Optional)

Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2

Measuring and alarm range

ART:0 ~ 300 mmHg

PA:-6 ~ 120 mmHg

CVP/RAP/LAP/ICP: -10 ~ 40 mmHg

P1/P2: -10 ~ 300 mmHg

Press Sensor

Sensitivity :5 uV/V/mmHg

Impedance: 300-3000Ω

Resolution: 1 mmHg 

Accuracy : ± 2% or 1mmHg, which great

Actualization interval: about 1 Sec.

 

ETCO2(Optional)

 

Measure range: 0% - 13%

Resolution: 1 mmHg

Accuracy : ±2 mmHg (< 5.0% Measurement Value)

±10%( > 5% Measurement Value)

Response Time: 180ms

 

 

CE Certificate:

professional veterinary patient monitor, veterinary equipment

About Us

Shenzhen Jumper Medical Equipment Company is a manufacture of medical equipment in China since 1998, with CE&FDA approved.

We have over 300 employees, cover an area of 8,000 M2

We are in the field of

-Fetal Doppler

-Fetal Monitor

-Patient Monitor

-Finger Pulse Oximeter

-Infrared Forehead Thermometer.

-Finger Pulse Oximeter.

professional veterinary patient monitor, veterinary equipment

professional veterinary patient monitor, veterinary equipment

professional veterinary patient monitor, veterinary equipment

If you feel good about our product, why don't you send an enquiry?

We promise to give you a reply in 24 hours.

If you did not find what you want,  Pls click here.

Send Inquiry to This Supplier

Message
0/5000

Want the best price? Post an RFQ now!
Business Type
Manufacturer
Year Established
1998
Factory Size
1,000-3,000 square meters
Total Employees
101 - 200 People

Recommended Categories