Home > Health & Medicine > IVD > Drawell-Crystal Auto Biochemistry Analyzer (300 T/H)

Drawell-Crystal Auto Biochemistry Analyzer (300 T/H)

Drawell-Crystal Auto Biochemistry Analyzer (300 T/H) photo-1
US$ 3,000 - 12,000 MOQ: 1 Piece
Key Specifications
Get Latest Price
Type:
Biochemical Analysis System, discrete, fully automatic
Brand Name:
DRAWELL
Model Number:
Drawell-CRYSTAL
Payment & Shipping
Payment Methods:
Port of Shipment:
Shanghai
Delivery Detail:
Delivery time depends on order quantity.
Brand Name DRAWELL
Model Number Drawell-CRYSTAL
Place of Origin Shanghai, China

Drawell-Crystal Auto Biochemistry Analyzer (300 T/H) 

 

1. Fully automatic, Discrete, Random access 

2.300 test per hour, 45 reagent position, 60 sample position and 120cuvettes. 

3. Refrigerated reagent tray, 24 hour working. 

4. Reversed optics and imported photo sensor, ensuring the precision of detection results. 

5. Precision dispense pumps assures the accuracy of adding sample volmue, sample volume step by 0.1uL, reagent volume step by 1uL. 

 

Overall

Performance

Model

Drawell-CRYSTAL

 

Type

discrete, fully automatic

 

Throughput

300T/H

 

Analysis Method

End-point, Fixed-time,Rate

 

Programming Items

programmable for any test items

 

Certification

CE, FDA, ISO9001, ISO14001, ISO13485

 

Sample unit

Sample Settings

60 sample positions, including calibration, quality control and emergency, primary tubes and serum is available

 

Sample Volume

2ul ~50ul (0.1ul stepping)

 

Sample Probe

Having liquid level induction function, ensuring the accuracy of adding sample

 

Probe cleaning

Automatic washing both interior and exterior

 

Analysis Items

analyzing maximally 45 items at the same time

 

Reagent unit

Reagent Position

45 low-temperature refrigerated reagent position

 

Reagent Volume

R1:25~300ul, R2:10~150ul (1ul stepping), supporting reagent from different countries

 

Temperature Control

continuous 2C~8Ccooling within 24 hours

 

Reagent Probe

Having liquid level induction function, ensuring the accuracy of adding reagent. Inventory checking

 

Probe cleaning

Automatic washing both interior and exterior

Reaction unit

Reaction Time

10 minutes

Temperature Control

37C±0.1C

 

Cuvettes

120 independent reaction cuvettes (optical path length 6mm)

 

Stirrer

after injecting reagent and sample, blending immediately

 

Cleaning

automatic water cleaning and drying system

 

System Adjustment

System can record the blank value and get more accurate result by deducting blank base value of each cuvette

Optical System

Light Source

12V/30W halogen-tungsten lamp, 2000 hours

 

Receiver

8 precision photo receivers

 

Wavelength Range

340,405,450,510,546,578,630,700mm

Calibration and QC

Calibration

one-point,multi-point

Linear Range

0.000~2.500Abs

 

Quality Control

Inserting quality control at random.6 quality controls can be reached with one item

 

Software

Emergency Function

emergency sample can be placed in the sample tray at any time

 

Software

Windows 2000 or Windows XP or above, sample and easy to operate

 

Data Processing

Having reaction curve detection and data display function in the whole process

Others

Power Supply

AC220V+22V 50Hz+1Hz

 

Dimension

1020mm×830mm×1200mm (W*D*H)

 

Weight

150kg

 

 

  

Drawell-Crystal Auto Biochemistry Analyzer (300 T/H)Drawell-Crystal Auto Biochemistry Analyzer (300 T/H)Drawell-Crystal Auto Biochemistry Analyzer (300 T/H)Drawell-Crystal Auto Biochemistry Analyzer (300 T/H)

 

Send Inquiry to This Supplier

Message
0/5000
Want the best price? Post an RFQ now!
Business Type
Manufacturer, Trading Company
Year Established
2014
Factory Size
1,000-3,000 square meters
Total Employees
101 - 200 People